Call us: 86-731-84012074

Contact us now for your demand[email protected]

Alternative Fuel

Nitrogen Determination in Biofuel (Sugaracane Bagasse& Pelletized Oil Palm Fiber)

Sample Preparation
Crush the sample to size <1mm

Instrument
5E-TCN2200 Nitrogen Analyzer

Instrument Settings
Oxygen Flow(SV4): 3L/min;
Oxygen Flow(SV5): 0.8L/min;
Oxygen Flow(SV6): 0.3L/min
Main Furnace Temperature: 950 ℃
Secondly Furnace Temperature: 850 ℃
Calibration Standard: GB/T 30728, UNI 15104, ISO 16948 and ASTM

Nominal Sample Weight: 80 mg

Analysis Time: 5 minutes

Special Modifications to Unit
None

Method Approval
GB/T 30728,UNI 15104,ISO 16948 and ASTM

Procedure
1. Prepare instrument as outlined in the operator’s instruction manual (perform maintenance and
leak checks, etc.).
2. Purging and analyze blanks until instrument is stable (nitrogen results will reach a plateau),
3. (4-5) blanks and (1-2) waste samples to condition the instrument
4. 3 samples of calibration substance to check or carry out the calibration;
5. Test 9 samples
6. 2 samples of calibration substance to check the calibration;
7. Repetition of point 5 to point 6 until all samples are processed;

Note: Do Calibrating refer to operator’s instruction manual for details.

Typical Results
Code Name Nad Range Average Mad(%) Nd(%)
2016062803 pelletized oil palm fiber 0.53 0.05 0.50 9.38 0.55
2016062803 pelletized oil palm fiber 0.48
2016062804 sugaracane Bagasse 0.42 0.06 0.45 7.42 0.49
2016062804 sugaracane Bagasse
0.48

customize your solutions request a quote right now